• Features :
    • DC 40KV CAT II Pollution 2
    • 600MΩ Impedance
    • 25KV ± 2%, 40KV ± 3%
    • T.C. 200 ppm/oC
    • Positive Polarity only.
    • Resolution 1KV
    • Need not power.
    • CE、IEC1010
        

    SPECIFICATIONS ( 20 ± 5℃, RH < 80% )
    ! WARNING !
    ?Before taking any measurements, first connect the alligator clip of this probe to earth ground and make sure connection is electrically good.
    Input impedance:600MΩ
    Maximum operation voltage:
    40KV DC , CAT II , Pollution 2
    Polarity:
    Positive only
    Display:
    Analog indicate
    Accuracy:
    Factory calibrater 25KV ± 2%
    ?????????FULL scale 40KV ± 3 %
    Temperature Coefficient:
    ≦ 200ppm / ℃
    Maximum loading Current:
    ≦ 70μA
    Maximum loading power:
    ≦ 3 watt
    Voltage Range:
    40KV / 1 Range
    Voltage resolution:
    1KV
    Power source:
    Need not power
    Operating temperature:
    0 ℃ ~ + 50 ℃
    Storage temperature:
    - 20 ℃ ~ + 70 ℃
    Ground lead length:
    90cm ( 35 " )
    Dimensions:
    420 L x 80ψ
    Weight :
    360g